Simplified efficiency calibration methods for semiconductor detectors used in criticality dosimetry

期刊

APPLIED RADIATION AND ISOTOPES
卷 187, 期 -, 页码 -

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2022.110335

关键词

Germanium detectors; Simplified efficiency calibration methods; True coincidence summing effect; Co-60; Na-24; Emergency planning; Neutron dosimetry; Gamma radiation/spectrometry

向作者/读者索取更多资源

Simplified methods for determining the photo-peak efficiency of semiconductor HPGe detector system are proposed and verified. These methods can be used to estimate neutron dose from irradiated human blood in potential criticality events.
Oversimplified and simplified methods based on true coincidence summing effect used in uncomplicated determination of the photo-peak efficiency of the semiconductor High Purity Germanium (HPGe) detector system are suggested and verified. The methods and calibrated Co-60 radioactive source may be used to commission any HPGe detector to use during potential criticality event. The determined accuracy of the semiconductor HPGe detector system using this method is a few percent (for the detector system used in this study it was similar or equal to 8% for oversimplified and similar or equal to 5% for simplified methods accordingly) reasonable, expected, and good enough to use for estimation of neutron dose from irradiated human blood in a potential criticality event. Moreover, if one can experimentally deduce the photo-peak efficiency for Co-60 1333 keV gamma-ray line using the suggested methods, then with a few percent accuracy this efficiency could be also used for 1369 keV gamma-ray line in the decay of Na-24 isotope.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据