Analytical Sensitivity of Eight Different SARS-CoV-2 Antigen-Detecting Rapid Tests for Omicron-BA.1 Variant

期刊

MICROBIOLOGY SPECTRUM
卷 10, 期 4, 页码 -

出版社

AMER SOC MICROBIOLOGY
DOI: 10.1128/spectrum.00853-22

关键词

Omicron-BA.1 variant; antigen-detecting rapid diagnostic tests; COVID-19; Omicron variant; SARS-CoV-2; variants of concern

资金

  1. Swiss National Science Foundation [196383, 196644]
  2. Fondation Privee des Hopiteaux Universitaires de Geneve
  3. FIND
  4. global alliance for diagnostics

向作者/读者索取更多资源

The sensitivity for detecting Omicron-BA.1 varies among different Ag-RDTs, highlighting the need for careful consideration when using these tests. Analytical and retrospective testing demonstrated variable sensitivity for both Omicron-BA.1 and Delta, but all seven Ag-RDTs showed comparable sensitivities in the retrospective study.
The emergence of each novel SARS-CoV-2 variant of concern (VOC) requires investigation of its potential impact on the performance of diagnostic tests in use, including antigen-detecting rapid diagnostic tests (Ag-RDTs). Although anecdotal reports have been circulating that the newly emerged Omicron-BA.1 variant is in principle detectable by Ag-RDTs, few data on sensitivity are available. We have performed (i) analytical sensitivity testing with cultured virus in eight Ag-RDTs and (ii) retrospective testing in duplicates with clinical samples from vaccinated individuals with Omicron-BA.1 (n = 59) or Delta (n = 54) breakthrough infection on seven Ag-RDTs. Overall, in our analytical study we have found heterogenicity between Ag-RDTs for detecting Omicron-BA.1. When using cultured virus, we observed a trend toward lower endpoint sensitivity for Omicron-BA.1 detection than for earlier circulating SARS-CoV-2 and the other VOCs. In our retrospective study, the detection of Delta and Omicron-BA.1 was assessed in a comparable set of stored clinical samples using seven Ag- RDTs. Four hundred ninety-seven of all 826 tests (60.17%) performed on Omicron-BA. 1 samples were positive, compared to 489/756 (64.68%) for Delta samples. In the analytical study, the sensitivity for both Omicron-BA.1 and Delta between the Ag-RDTs was variable. All seven Ag-RDTs showed comparable sensitivities to detect Omicron-BA.1 and Delta in the retrospective study. IMPORTANCE Sensitivity for detecting Omicron-BA.1 shows high heterogenicity between Ag-RDTs, necessitating a careful consideration when using these tests to guide infection prevention measures. Analytical and retrospective testing is a proxy and timely solution to generate rapid performance data, but it is not a replacement for clinical evaluations, which are urgently needed. Biological and technical reasons for detection failure by some Ag-RDTs need to be further investigated.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据