Reproducibility of apparent diffusion coefficient measurement in normal prostate peripheral zone at 1.5T MRI

期刊

DIAGNOSTIC AND INTERVENTIONAL IMAGING
卷 103, 期 11, 页码 545-554

出版社

ELSEVIER MASSON, CORP OFF
DOI: 10.1016/j.diii.2022.06.001

关键词

Prostate; Diffusion magnetic resonance imaging; Healthy volunteers; Reproducibility of results

向作者/读者索取更多资源

The purpose of this study was to quantify the influence of factors of variability on ADC estimation in the normal prostate peripheral zone. The results showed that the MRI scanner, the lowest b-value used, and the period of the day had a substantial impact on ADC computation.
Purpose: The purpose of this study was to quantify the influence of factors of variability on apparent diffusion coefficient (ADC) estimation in the normal prostate peripheral zone (PZ).Materials and methods: Fifty healthy volunteers underwent in 2017 (n = 17) or 2020 (n = 33) two-point (0, 800 s/mm2) prostate diffusion-weighted imaging in the morning on 1.5 T scanners A and B from different manufacturers. Additional five-point (50, 150, 300, 500, 800 s/mm2) acquisitions were performed on scanner B in the morning and evening. ADC was measured in PZ at midgland using ADC maps reconstructed with var-ious b-value combinations. ADC distributions from 2017 and 2020 were compared using Wilcoxon rank sum test. ADC obtained in the same volunteers were compared using Bland Altman methodology. The 95% confi-dence interval upper limit of the repeatability/reproducibility coefficient defined the lowest detectable ADC difference.Results: Forty-nine participants with a mean age of 24.6 +/- 3.8 [SD] years (range: 21-37 years) were finally included. ADC distributions from 2017 and 2020 were not significantly different and were combined. Despite high individual variability, there was no significant bias (10 x 10-6 mm2/s, P = 0.58) between ADC measure-ments made on both scanners. On scanner B, differences in lowest b -values chosen within the 0-500 s/mm2 range for two-point ADC computation induced significant biases (56-109 x 10-6 mm2/s, P < 0.0001). ADC was significantly lower in the morning (bias: 33 x 10-6 mm2/s, P = 0.006). The number of b-values had little influence on ADC values. The lowest detectable ADC difference varied from 85 x 10-6 to 311 x 10-6 mm2/s across scanners, b-value combinations and periods of the day.Conclusions: The MRI scanner, the lowest b-value used and the period of the day induce substantial variability in ADC computation.(c) 2022 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据