Investigation of the 176Yb Interference Correction during Determination of the 176Hf/177Hf Ratio by Laser Ablation and Solution Analysis on the Neoma MC-ICP-MS

期刊

MINERALS
卷 12, 期 7, 页码 -

出版社

MDPI
DOI: 10.3390/min12070882

关键词

zircon; Hf isotopes; MC-ICP-MS; laser ablation

资金

  1. US Department of Energy (DOE) [DEAC05-00OR22725]

向作者/读者索取更多资源

In this study, the behavior of the Lu-Yb-Hf system during laser ablation and solution-based analyses was assessed using the recently released Neoma (TM) platform. The primary focus was on characterizing the Yb interference correction, which is an important issue in Hf isotope analysis. The results showed that high signal intensity analyses produced accurate and precise data, while an empirical approach was effective in obtaining accurate results with low Yb signals.
We utilized the Neoma (TM), a recently released MC-ICP-MS platform offered by ThermoFisher Scientific, to assess the behavior of the Lu-Yb-Hf system during laser ablation analyses of various zircon standards as well as solution-based analyses of the JMC-475 Hf standard doped with varying quantities of Yb and Lu. The primary goal of this work was to characterize the behavior of the Yb interference correction on the Neoma (TM) platform since this is one of the biggest issues in the Hf isotope analysis community and because the Neoma (TM) platform will supplant the Neptune (TM) series instrument. During laser ablation analysis, we found that the overall data quality scales proportionally with the total Hf signal intensity, with higher signal analyses producing extremely accurate (within 1 epsilon(Hf) unit) and precise (sub epsilon(Hf) unit within-run standard errors) data. At low Yb signals (<0.1 V Yb-173), we were not able to produce an accurate internal Yb mass bias factor. However, utilizing an empirical approach allows for the application of session-specific relationships between the Yb and Hf mass bias factors, determined by analysis of standards of varying Yb content, to produce accurate epsilon(Hf) values from zircons with higher Yb/Hf ratios even where the total Hf signal intensity is relatively low. Similar behavior was observed in the solution analyses. Lastly, while the behavior of the Yb interference correction on the Neoma (TM) platform appears comparable to the Neptune (TM) series MC-ICP-MS, further work will help refine the understanding of the controls on mass bias behavior, oxide formation, session-to-session stability, etc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据