4.4 Article

Disintegration of the Top Stent on Zenith Abdominal Aortic Stent-Grafts

期刊

JOURNAL OF ENDOVASCULAR THERAPY
卷 23, 期 2, 页码 302-306

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/1526602815626559

关键词

abdominal aortic aneurysm; aneurysm growth; complication; device failure; endograft; endoleak; endovascular aneurysm repair; explantation; fenestrated cuff; migration; proximal fixation stent; stent-graft

向作者/读者索取更多资源

Purpose: To describe a heretofore unreported complication involving the Zenith Low Profile (LP) stent-graft. Case Report: Two men, aged 75 and 67 years, respectively, underwent abdominal aortic aneurysm repair with a Zenith LP device. At 4 and 3 years, respectively, computed tomography angiography revealed separation of the proximal fixation stent from the stent-graft. In the first patient, there was stent-graft migration but no evidence of an endoleak; however, the aneurysm had grown. A fenestrated cuff was placed, sealing distally in the previous LP graft. The second patient had a type I endoleak. Open surgery was performed, and the main body of the graft was explanted. Postoperative examination of the device revealed that the fixation sutures on the suprarenal stent were still attached to the stent and had eroded through the graft material. Conclusion: Physicians should be aware of the potential for top stent separation from the Zenith LP stent-graft as a cause of endoleak and migration.