Two Instrument Comparison of Reagents From a US FDA-Approved Assay for the Assessment of Ki-67 in High-Risk Early Breast Cancer

期刊

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/PAI.0000000000001050

关键词

abemaciclib; early breast cancer; Ki-67; MIB-1; Omnis

资金

  1. Eli Lilly and Company
  2. Agilent Technologies

向作者/读者索取更多资源

This study compared the concordance of Ki-67 immunohistochemistry results obtained from different automated staining instruments on breast cancer samples and found that good concordance can be achieved with reagents run on the ASL48 instrument, demonstrating the importance of optimized protocols and standardized scoring.
The objective of this study was to measure concordance of results obtained from the US Food and Drug Administration-approved Ki-67 immunohistochemistry MIB-1 pharmDx assay performed on the Dako Omnis automated staining instrument (Omnis) versus results produced from the assay reagents applied using an optimized protocol on the more widely available Autostainer Link 48 (ASL48) platform. Tissue sections obtained from 40 formalin-fixed paraffin-embedded breast carcinoma samples, with available Oncotype DX Breast Recurrence Score (RS) results, were stained. Three certified pathologists scored slides at 3 timepoints, totaling 360 observations for each instrument (N=720 total) using the approved scoring approach. Using the >= 20% cutoff, agreement was calculated with corresponding 2-sided 95% percentile bootstrap confidence intervals (CIs). Pairwise comparisons (N=360) from the interinstrument evaluation, performed with all observers, resulted in 325 (90.3%) concordant outcomes (244 negative and 81 positive) and 35 (9.7%) discordant outcomes. The overall agreement was 90.3% (95% confidence interval, 85.6% to 94.4%). No significant systematic differences were observed between instruments. Specimens scored from the Omnis were on average <1% higher than ASL48, with high correlation and little bias between the continuous Ki-67 scores (concordance correlation coefficient=0.916). Most specimens with a Ki-67 score >= 20% had a RS >25. This study demonstrated that good concordance can be achieved with the reagents run on the ASL48 instrument when using an optimized protocol and standardized scoring.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据