期刊
JOURNAL OF ELECTROANALYTICAL CHEMISTRY
卷 763, 期 -, 页码 37-44出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/j.jelechem.2015.12.039
关键词
Glucose sensors; Flower-like CuO/MWCNTs; Screen printing
Two non-enzymatic glucose sensors based on ordered self-assembly of flower-like CuO and CuO/MWCNTs modified graphite electrodes have been successfully fabricated by a facile and effective screen-printed way. The prepared flower-like CuO and CuO/MWCNTs were characterized by scanning electron microscopy (SEM), transmission electron microscopy (TEM) and X-ray diffraction (XRD). The cyclic voltammetry (CV) and chronoamperometry demonstrated that the CuO/MWCNTs modified electrode displayed much higher electro-catalytic activity than the CuO modified electrode toward glucose, exhibiting a high sensitivity of 1.211 mA mM(-1) cm(-2), a fast response time of 3 s, a low detection limit of 4 mu M (S / N = 3) and three linear regions from 4 mu M to 14.5 mM. The sensors can effectively resist the effect of interferences such as UA, DA, AA, fructose, lactose, sucrose and NaCl. In addition, the results tested by real blood serum samples showed that the optimized electrodes were applicable for determination of glucose in real blood serum samples. Thus, the CuO and CuO/MWCNTs modified electrodes are potential candidates for routine glucose analysis. (C) 2015 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐