4.6 Review

Advances in the Mass Sensitivity Distribution of Quartz Crystal Microbalances: A Review

期刊

SENSORS
卷 22, 期 14, 页码 -

出版社

MDPI
DOI: 10.3390/s22145112

关键词

quartz crystal microbalance (QCM); mass sensitivity; electrode parameter; particle displacement amplitude

资金

  1. National Natural Science Foundation of China [61871098]

向作者/读者索取更多资源

This article introduces the application of quartz crystal microbalance (QCM) and the importance of mass sensitivity. It focuses on the influence of electrode parameters and analyzes the effect of operating frequency on mass sensitivity.
A quartz crystal microbalance (QCM) is a typical acoustic transducer that undergoes a frequency shift due to changes in the mass of its surface. Its high sensitivity, robustness, small size design, and digital output have led to its widespread development for application in the fields of chemistry, physics, biology, medicine, and surface science. Mass sensitivity is one of the vital parameters and forms the basis for quantitative analysis using QCMs. This review firstly introduces the importance, definition, calculation, and measuring method of the mass sensitivity and then focuses on reviewing the influence of electrode parameters (including electrode shape, electrode diameter, electrode thickness, electrode material, etc.) on the mass sensitivity distribution of QCMs. Finally, the effect of the operating frequency on the mass sensitivity of QCMs is also analyzed.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据