This study developed a high-performance liquid chromatography method for simultaneous quantification of multiple cannabinoids in C. sativa using an AQbD-driven approach. The method showed satisfactory selectivity, linearity, precision, and ruggedness, making it suitable for quality control of cannabis-based products.
Introduction Medical uses of Cannabis sativa L. have gained interest in recent decades, which highlights the need for defining appropriate quality specifications for Cannabis-based products. However, the complexity of plant matrices and structural similarity between cannabinoids make analytical development a challenging task. Thus, the application of analytical quality by design (AQbD)-driven approaches can favour the development of fit-for-purpose methods. Objectives To develop a high-performance liquid chromatography diode array detector (HPLC-DAD) method for simultaneous quantification of cannabidiol, Delta(9)-tetrahydrocannabinol, cannabidiolic acid, tetrahydrocannabinolic acid, and cannabinol in C. sativa by applying an AQbD-driven approach. Materials and Methods Critical method attributes (CMA) were established following the analytical target profile. Critical method variables (CMV) were categorised based on risk assessment and literature review. Selected CMV regarding sample preparation and chromatographic conditions were optimised using response surface methodology (RSM). The working point was estimated by multiple response optimisation using Deringer's desirability function. The validity of the optimal conditions was confirmed experimentally. Method validation was performed according to ANVISA and ICH guidelines. Relative response factors (RRFs) were also determined. Results and Discussion Baseline resolution of 12 major cannabinoids was achieved in a 35 min chromatographic analysis. All experimental responses obtained during confirmatory analyses were within the prediction intervals (PI95%). Method's selectivity, linearity (10-100 mu g/mL), precision, bias, extraction recovery, and ruggedness were satisfactorily demonstrated. Conclusions The application of an AQbD-driven approach allowed for a better understanding of the effects of the ensemble of CMV on the analyte's behaviour, enabling the definition of appropriate conditions to ensure consistent achievement of the intended method's performance.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据