4.2 Article

Comparison of assessment of coagulation in healthy dogs by the TEG 6s and TEG 5000 viscoelastic analyzers

期刊

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/10406387221112638

关键词

canine; coagulation; coagulopathy; TEG 5000; TEG 6s; thromboelastography; viscoelastic

资金

  1. Department of Defense

向作者/读者索取更多资源

The TEG 6s point-of-care viscoelastic analyzer has potential for rapid viscoelastic analysis in veterinary medicine, but agreement with the TEG 5000 analyzers was poor, only showing satisfactory agreement for certain measurements in healthy dogs.
The TEG 6s (Haemonetics) point-of-care viscoelastic analyzer is portable, compact, simple to use, and has the potential for rapid viscoelastic analysis that can guide the treatment of veterinary patients at the site of care. Although approved for use in people, the TEG 6s has yet to be evaluated for hemostatic analysis in veterinary medicine. Citrated whole blood (CWB) was collected from 27 healthy dogs. An aliquot of CWB from each dog was diluted by 33% with an isotonic crystalloid, representing an in vitro model of hemodilution. Unaltered and diluted CWB samples were analyzed using 2 TEG 6s and 6 TEG 5000 (Haemonetics) analyzers. The 6 TEG 5000 analyzers ran duplicate analyses of either unaltered or diluted samples using 1 of 3 reagents (Haemonetics): Kaolin TEG, RapidTEG, or TEG Functional Fibrinogen. Duplicate TEG 5000 analyses were averaged and compared with a single TEG 6s analysis. Lin concordance correlation coefficient and Bland-Altman plots were used to evaluate agreement of reaction time, kinetic time, alpha angle, maximum amplitude (MA), and G value (G) for samples activated with Kaolin TEG, and agreement of MA for samples activated with RapidTEG between the 2 machines. Overall, agreement between the TEG 6s and TEG 5000 analyzers was poor. Viscoelastic measurements by the TEG 6s and TEG 5000 in healthy dogs were not all interchangeable. Agreement was satisfactory only for MA and G measurements of diluted blood samples activated with Kaolin TEG, and MA measurements for both unaltered and diluted blood samples activated with RapidTEG.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据