Application of Tele-Ultrasound Robot During COVID-19 Pandemic A Feasibility Study

期刊

JOURNAL OF ULTRASOUND IN MEDICINE
卷 42, 期 3, 页码 595-601

出版社

WILEY
DOI: 10.1002/jum.16041

关键词

coronavirus disease 2019; novel coronavirus pneumonia; tele-ultrasound; ultrasonic robot

向作者/读者索取更多资源

The study aimed to investigate the accuracy of ultrasonic diagnosis using the tele-ultrasound robot in Leishen Shan Hospital. The results showed that the tele-ultrasound robot method had comparable accuracy to the conventional method in diagnosing common diseases in COVID-19 patients.
Objective To investigate the accuracy of ultrasonic diagnosis using the tele-ultrasound robot in Leishen Shan Hospital. Method Twenty-two patients with novel coronavirus pneumonia from Leishen Shan Hospital voluntarily participated in this study. Their thyroids, neck vessels, hepatobiliaries and kidneys were scanned by both tele-ultrasound robot manufactured by Imabot Co., Ltd, Wuhan and conventional method. The ultrasound diagnosis of each patient was compared, and the ultrasound images obtained by the two methods were mixed together and double-blindly diagnosed by an experienced ultrasound radiologist. Results There were 44 positive lesions in 110 sites of 22 patients. Of which the two methods, 40 positive lesions were detected by the robotic method with 4 lesions missed (2 small polyps of gallbladder, 1 small hemangioma of liver and 1 small cyst of kidney) and 1 lesion misdiagnosed (normal carotid artery was misdiagnosed as carotid atherosclerotic plaque); 44 positive lesions were detected by conventional method with 1 small cyst of the liver was missed. There was no statistically significant difference in the accuracy rate between the robotic method and the conventional method using the chi-square test of the four-grid data (P>.05). Conclusion The application of tele-ultrasound robot meets the standard of patient care during the pandemic. The method is feasible to provide adequate ultrasound information to diagnose common abdominal, vascular, superficial organ pathologies in patients with COVID-19 with acceptable accuracy compared with a conventional ultrasound scan.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据