Impact of systematic MLC positional uncertainties on the quality of single-isocenter multi-target VMAT-SRS treatment plans

期刊

出版社

WILEY
DOI: 10.1002/acm2.13708

关键词

brain metastases; leaf offset; leaf positional uncertainty; MLC; single isocenter; spatial accuracy; SRS; stereotactic radiosurgery; VMAT

资金

  1. General Secretariat for Research and Technology (GSRT), under the HFRI PhD Fellowship grant [74112/2017]

向作者/读者索取更多资源

The impact of systematic MLC leaf positional uncertainties on the quality of intracranial single-isocenter multi-target VMAT-SRS treatment plans is studied in this research. The study attempts to estimate appropriate tolerance levels and finds that plan quality indices and dose-volume metrics are very sensitive to systematic sub-millimeter leaf positional inaccuracies. Tailoring acceptable and tolerance levels to VMAT-SRS spatial and dosimetric accuracy requirements is necessary.
Purpose To study the impact of systematic MLC leaf positional uncertainties (stemming from mechanical inaccuracies or sub-optimal MLC modeling) on the quality of intracranial single-isocenter multi-target VMAT-SRS treatment plans. An estimation of appropriate tolerance levels is attempted. Methods Five patients, with three to four metastases and at least one target lying in close proximity to organs-at-risk (OARs) were included in this study. A single-isocenter multi-arc VMAT plan per patient was prepared, which served as the reference for dosimetric impact evaluation. A range of leaf offsets was introduced (+/- 0.03 mm up to +/- 0.30 mm defined at the MLC plane) to both leaf banks, by varying the leaf offset MLC modeling parameter in Monaco for all the prepared plans, in order to simulate projected leaf offsets of +/- 0.09 mm up to +/- 0.94 mm at the isocenter plane, respectively. For all offsets simulated and cases studied, dose distributions were re-calculated and compared with the corresponding reference ones. An experimental dosimetric procedure using the SRS mapCHECK diode array was also performed to support the simulation study results and investigate its suitability to detect small systematic leaf positional errors. Results Projected leaf offsets of +/- 0.09 mm were well-tolerated with respect to both target dosimetry and OAR-sparing. A linear relationship was found between D-95% percentage change and projected leaf offset (slope: 12%/mm). Impact of projected offset on target dosimetry was strongly associated with target volume. In two cases, plans that could be considered potentially clinically unacceptable (i.e., clinical dose constraint violation) were obtained even for projected offsets as small as 0.19 mm. The performed experimental dosimetry check can detect potential small systematic leaf errors. Conclusions Plan quality indices and dose-volume metrics are very sensitive to systematic sub-millimeter leaf positional inaccuracies, projected at the isocenter plane. Acceptable and tolerance levels in systematic MLC uncertainties need to be tailored to VMAT-SRS spatial and dosimetric accuracy requirements.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据