4.6 Article

Design and evaluation of microfluidic devices for two-dimensional spatial separations

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1434, 期 -, 页码 127-135

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2016.01.003

关键词

Microfluidic devices; Spatial separation systems; Two-dimensional liquid chromatography; Computational fluid dynamics

资金

  1. EUREKA organization [C.2332.0227]
  2. Thermo Fisher Scientific
  3. Research Foundation Flanders (FWO-Vlaanderen)
  4. Institute for promotion of innovation through science and technology in Flanders (IWT-Flanders)

向作者/读者索取更多资源

Various designs of chips for comprehensive two-dimensional spatial liquid chromatography were investigated. The performance of these chips was initially evaluated using computational fluid dynamics (CFD). A bifurcating distributor with an angle of 140 between branches was implemented in order to achieve a homogeneous velocity field. The cross-sectional area of the channels of the flow distributor was fixed at 0.5 x 0.5 mm, which allows a robust micromilling technique to be used for chip manufacturing. Experiments were performed with chips featuring purposely introduced imperfections in the structure of the bifurcating flow distributor to study its capacity of overcoming potential local clogging. Split peaks were observed when 75% of one of the flow channels was obstructed, in line with the CFD predictions. The main bottlenecks for the performance of the spatial two-dimensional chips were identified, viz. sample injected in the first dimension diverging into the flow distributor and channel discretization (i.e., remixing of first-dimension separation peaks because of finite number of second-dimension channels). Solutions to the former problem were studied by applying a flow resistance in the vertical segments that formed the outlets of the flow distributor and by simulating the presence of constrictions. It was found that a flow resistance of 1.0 x 10(11) m(-2) reduced the amount of sample diverging into the flow distributor by a factor of 10. The presence of a constriction of 90% of the segment area and 50% of the segment length decreased the diverging flow by a factor of 5. The influence of the linear velocity was significant. Solutions to the channel discretization problem were sought by investigating different designs of spatial two-dimensional chips. (C) 2016 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据