4.7 Article

Evaluation of Closed System Transfer Devices in Preventing Chemotherapy Agents Contamination During Compounding Process-A Single and Comparative Study in China

期刊

FRONTIERS IN PUBLIC HEALTH
卷 10, 期 -, 页码 -

出版社

FRONTIERS MEDIA SA
DOI: 10.3389/fpubh.2022.827835

关键词

CSTD; pharmacy; chemotherapy agents; closed system transfer devices; occupational protection

向作者/读者索取更多资源

This study compared the efficacy of closed system transfer devices (CSTDs) with traditional devices in the safe handling of injectable hazardous drugs. The results showed that CSTDs significantly reduced drug contamination and improved pharmacist satisfaction and safety perception.
AimWe performed a comparative study to investigate the efficacy of closed system transfer devices (CSTDs) on the safe handling of injectable hazardous drugs (HDs). MethodsThe exposure assessments of cyclophosphamide and cytarabine were performed under traditional or CSTDs. For preparation activity, chemotherapy contamination samples on protective equipment (such as gloves and masks) were collected. The contamination analysis was performed by liquid chromatography with tandem mass spectrometry (LC-MS/MS). A 6-item form was distributed monthly (form M1-M6, total 6 months) to assess the pharmacists' experience on ergonomics, encumbrance, and safety impression. ResultsTotally, 96 wiping samples were collected throughout the study. The numbers of contaminated cyclophosphamide samples reduced under CSTD were -37.8, -41.6, -67.7, -47.3, and -22.9% and cytarabine were -12.3, -12.1, -20.6, -69.6, and -56.7% for left countertop, right countertop, medial glass, air-intake vent and door handle, as compared to traditional devices. The reduction was similar to pharmacist devices, i.e., -48.2 and -50.0% for masks and gloves cyclophosphamide contamination, -18.0 and -42.4% for cytarabine. This novel system could improve contamination on dispensing table, transfer container, and dispensing basket by -16.6, -6.0, and -22.3% for cyclophosphamide and -28.5, -22.5, and -46.2% for cytarabine. A high level of satisfaction was consistently associated with ergonomics for CSTD during the compounding process. Meanwhile, a slightly decreased satisfaction on ergonomics, encumbrance, and safety impression was observed for the traditional system between M2 and M3. ConclusionClosed system transfer devices are offering progressively more effective alternatives to traditional ones and consequently decrease chemotherapy exposure risk on isolator surfaces.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据