4.6 Article

Diagnostic Performance of Xpert MTB/RIF Ultra Compared with Predecessor Test, Xpert MTB/RIF, in a Low TB Incidence Setting: a Retrospective Service Evaluation

期刊

MICROBIOLOGY SPECTRUM
卷 10, 期 3, 页码 -

出版社

AMER SOC MICROBIOLOGY
DOI: 10.1128/spectrum.02345-21

关键词

Ireland; Mycobacterium tuberculosis; PCR; TB; Xpert MTB; RIF; Xpert MTB; RIF Ultra; diagnostic performance

资金

  1. Irish Mycobacteria Reference Laboratory, St James's Hospital
  2. Department of Clinical Microbiology, Trinity College Dublin

向作者/读者索取更多资源

The study demonstrates the improved sensitivity of the Ultra compared with the Xpert, particularly in smear negative TB disease, for both pulmonary and extrapulmonary samples in a low TB incidence setting. Cycle threshold (Ct) value for both assays was found to positively correlate with time to TB culture positivity, suggesting that Ct and semiquantitative results could be used as indicators of sample MTBC bacillary burden, and thus, perhaps, of transmission potential. This may have implications for the designation of patient isolation precautions.
The aim of this study was to evaluate the performance of Xpert MTB/RIF Ultra (Ultra) compared with its predecessor, Xpert MTB/RIF (Xpert), in the diagnosis of tuberculosis (TB) in a low TB incidence country. Retrospective analysis was performed on 689 clinical samples received between 2015 and 2018, on which Xpert was performed, and on 715 samples, received between 2018 and 2020, on which Ultra was performed. Samples were pulmonary (n = 830) and extrapulmonary (n = 574) in nature, and a total of 264 were culture positive for Mycobacterium tuberculosis complex (MTBC). The diagnostic performance of both assays was analyzed using culture as the reference standard. The sensitivity of Ultra for culture positive (smear positive and smear negative) MTBC samples, was 93.2% (110/118) compared with 82.2% (120/146) for Xpert (P = 0.0078). In smear negative-culture positive samples, Ultra had a sensitivity of 74.2% (23/31) versus 36.11% (13/36) for Xpert (P = 0.0018). Specificity of both assays was comparable at 94.8% (566/597) for Ultra and 95.8% (520/543) for Xpert (P = 0.4475). The sensitivity of Ultra and Xpert assays among exclusively pulmonary samples was 95.3% (82/86) and 90.3% (84/93), respectively (P = 0.1955), and 87.5% (28/32) and 67.9% (36/53), respectively, among extrapulmonary samples (P = 0.0426). Ultra showed improved performance compared with Xpert in a low TB incidence setting, particularly in smear negative and extrapulmonary MTBC disease. The specificity of Ultra was lower than Xpert, however, this was not statistically significant. IMPORTANCE The study demonstrates the improved sensitivity of the Ultra compared with the Xpert, particularly in smear negative TB disease, for both pulmonary and extrapulmonary samples in a low TB incidence setting. Cycle threshold (Ct) value for both assays was found to positively correlate with time to TB culture positivity, suggesting that Ct and semiquantitative results could be used as indicators of sample MTBC bacillary burden, and thus, perhaps, of transmission potential. This may have implications for the designation of patient isolation precautions. The study demonstrates the improved sensitivity of the Ultra compared with the Xpert, particularly in smear negative TB disease, for both pulmonary and extrapulmonary samples in a low TB incidence setting. Cycle threshold (Ct) value for both assays was found to positively correlate with time to TB culture positivity, suggesting that Ct and semiquantitative results could be used as indicators of sample MTBC bacillary burden, and thus, perhaps, of transmission potential.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据