Influence of Artificial Soft Tissue on Intra-Operative Vibration Analysis Method for Primary Fixation Monitoring in Cementless Total Hip Arthroplasty

期刊

APPLIED SCIENCES-BASEL
卷 12, 期 8, 页码 -

出版社

MDPI
DOI: 10.3390/app12084027

关键词

Total Hip Arthroplasty; primary fixation; femoral stem fixation; smart instrumentation; vibration analysis; implant fixation monitoring

资金

  1. Internal Funds KU Leuven

向作者/读者索取更多资源

This study presents a dynamic response measurement protocol for determining the insertion endpoint of an implant in cementless total hip arthroplasty, and evaluates this protocol in the presence of artificial soft tissue. The results show that the technique can produce higher fixation measurements earlier in the insertion process when artificial soft tissue is present, but the soft tissue limits the system's ability to detect fixation changes.
Featured Application A decision supporting monitoring technique that informs the orthopedic surgeon when the femoral implant insertion endpoint is reached during cementless total hip arthroplasty. In cementless Total Hip Arthroplasty (THA), achieving high primary implant fixation is crucial for the long-term survivorship of the femoral stem. While orthopedic surgeons traditionally assess fixation based on their subjective judgement, novel vibration-analysis fixation-monitoring techniques show promising potential in providing the surgeon with objective and quantifiable fixation measurements. This study presents a dynamic response measurement protocol for implant endpoint insertion and evaluates this protocol in the presence of artificial soft tissue. After the artificial femur was prepared in accordance with the THA protocol, the implant was inserted and progressively hammered into the cavity. The Pearson Correlation Coefficient (PCC) and Frequency Response Assurance Criterion (FRAC) corresponding to each insertion hammer hit were derived from the Frequency Response Functions (FRF) corresponding to each insertion step. The protocol was repeated with the artificial femur submerged in artificial soft tissue to imitate the influence of anatomical soft tissue. The FRAC appeared overall more sensitive than the PCC. In the presence of the artificial soft tissue the technique yielded higher PCC and FRAC values earlier in the insertion process. The measurements with artificial soft tissue produced FRFs with fewer peaks, lower resonance frequencies, and overall higher damping factors. The soft tissue appears to limit the fixation-change detection capabilities of the system and a promising potential remedy to this limitation is suggested.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据