Gattinoni's equation is commonly used to calculate the mechanical power of ventilation, but it only calculates inspiratory power. Including PEEP in the equation is debated because PEEP does not contribute to power. Measuring the area within the pressure-volume loop accurately reflects the power received in a ventilation cycle and is not influenced by PEEP.
Gattinoni's equation, MPrs = 0.098 x RR x [V-T(2) (1/2E(rs) + RR x 1 + I:E/60 x I:E x R-aw) + V-T x PEEP], now commonly used to calculate the mechanical power (MP) of ventilation. However, it calculates only inspiratory MP. In addition, the inclusion of PEEP in Gattinoni's equation raises debate because PEEP does not produce net displacement or contribute to MP. Measuring the area within the pressure-volume loop accurately reflects the MP received in a whole ventilation cycle and the MP thus obtained is not influenced by PEEP. The MP of 25 invasively ventilated patients were calculated by Gattinoni's equation and measured by integration of the areas within the pressure-volume loops of the ventilation cycles. The MP obtained from both methods were compared. The effects of PEEPs on MP were also evaluated. We found that the MP obtained from both methods were correlated by R-2 =0.75 and 0.66 at PEEP 5 and 10 cmH(2)O, respectively. The biases of the two methods were 3.13 (2.03 to 4.23) J/min (P < 0.0001) and - 1.23 (- 2.22 to - 0.24) J/min (P =0.02) at PEEP 5 and 10 cmH(2)O, respectively. These P values suggested that both methods were significantly incongruent. When the tidal volume used was 6 ml/Kg, the MP by Gattinoni's equation at PEEP 5 and 10 cmH(2)O were significantly different (4.51 vs 7.21 J/min, P < 0.001), but the MP by PV loop area was not influenced by PEEPs (6.46 vs 6.47 J/min, P=0.331). Similar results were observed across all tidal volumes. We conclude that the Gattinoni's equation is not accurate in calculating the MP of a whole ventilatory cycle and is significantly influenced by PEEP, which theoretically does not contribute to MP.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据