Fibre Bragg Grating Based Interface Pressure Sensor for Compression Therapy

期刊

SENSORS
卷 22, 期 5, 页码 -

出版社

MDPI
DOI: 10.3390/s22051798

关键词

optical fibre sensor; compression therapy; venous leg ulcer; sub-bandage pressure sensor; Fibre Bragg Grating

资金

  1. EPSRC Impact Acceleration Award [EP/K503800/1]
  2. Karl Otto Braun GmbH
  3. Footfalls and Heartbeats (UK) Ltd.
  4. EPSRC
  5. Royal Society

向作者/读者索取更多资源

This paper presents the development of a fiber optic pressure sensor for measuring sub-bandage pressure, which eliminates the interference caused by fiber strain and temperature without affecting patient comfort. The sensor's reliability and accuracy are validated by comparisons with the current gold standard medical sensor.
Compression therapy is widely used as the gold standard for management of chronic venous insufficiency and venous leg ulcers, and the amount of pressure applied during the compression therapy is crucial in supporting healing. A fibre optic pressure sensor using Fibre Bragg Gratings (FBGs) is developed in this paper to measure sub-bandage pressure whilst removing cross-sensitivity due to strain in the fibre and temperature. The interface pressure is measured by an FBG encapsulated in a polymer and housed in a textile to minimise discomfort for the patient. The repeatability of a manual fabrication process is investigated by fabricating and calibrating ten sensors. A customized calibration setup consisting of a programmable translation stage and a weighing scale gives sensitivities in the range 0.4-1.5 pm/mmHg (2.6-11.3 pm/kPa). An alternative calibration method using a rigid plastic cylinder and a blood pressure cuff is also demonstrated. Investigations are performed with the sensor under a compression bandage on a phantom leg to test the response of the sensor to changing pressures in static situations. Measurements are taken on a human subject to demonstrate changes in interface pressure under a compression bandage during motion to mimic a clinical application. These results are compared to the current gold standard medical sensor using a Bland-Altman analysis, with a median bias ranging from -4.6 to -20.4 mmHg, upper limit of agreement (LOA) from -13.5 to 2.7 mmHg and lower LOA from -32.4 to -7.7 mmHg. The sensor has the potential to be used as a training tool for nurses and can be left in situ to monitor bandage pressure during compression therapy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据