4.6 Article

A Kalman Filter Approach for Estimating Tendon Wave Speed from Skin-Mounted Accelerometers

期刊

SENSORS
卷 22, 期 6, 页码 -

出版社

MDPI
DOI: 10.3390/s22062283

关键词

Kalman filter; shear wave tensiometry; tendon mechanics

资金

  1. National Science Foundation [NSF GRFP DGE-1747503, CBET-2019621]
  2. National Institutes of Health [NIH NIAMS F32-AR076267, R42-AR074897]

向作者/读者索取更多资源

This study investigates the use of a Kalman filter and redundant accelerometers to improve the accuracy and robustness of shear wave tensiometry in measuring tendon wave speed. The results show that the Kalman filter significantly improves the accuracy of estimated wave speeds, and increasing the number of accelerometers and using redundant accelerometers further reduce wave speed errors and improve the robustness of wave speed measures.
Shear wave tensiometry is a noninvasive approach for assessing in vivo tendon forces based on the speed of a propagating shear wave. Wave speed is measured by impulsively exciting a shear wave in a tendon and then assessing the wave travel time between skin-mounted accelerometers. Signal distortion with wave travel can cause errors in the estimated wave travel time. In this study, we investigated the use of a Kalman filter to fuse spatial and temporal accelerometer measurements of wave propagation. Spatial measurements consist of estimated wave travel times between accelerometers. Temporal measurements are the change in wave arrival at a fixed accelerometer between successive impulsive taps. The Kalman filter substantially improved the accuracy of estimated wave speeds when applied to simulated tensiometer data. The variability of estimated wave speed was reduced by similar to 55% in the presence of random sensor noise. It was found that increasing the number of accelerometers from two to three further reduced wave speed errors by 45%. The use of redundant accelerometers (>2) also improved the robustness of wave speed measures in the presence of uncertainty in accelerometer location. We conclude that the use of a Kalman filter and redundant accelerometers can enhance the fidelity of using shear wave tensiometers to track tendon wave speed and loading during movement.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据