Heavy-water-based moderator design for an AB-BNCT unit using a topology optimization algorithm

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 67, 期 10, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1361-6560/ac6723

关键词

boron neutron capture therapy; neutron moderator; topology optimization

向作者/读者索取更多资源

This study proposes a method using topology optimization algorithm and advanced Monte-Carlo transport code to optimize neutron moderator design. Through the application of this method, the generated neutron moderators have compact structures and successfully limit the exposure of patient's healthy tissues to levels below recommended limits.
Objective. The design of neutron moderators for BNCT treatment units currently relies on parametric approaches, which yield quality results but are ultimately limited by human imagination. Efficient but non-intuitive design solutions may thus be missed out. This limitation needs to be addressed. Approach. To overcome this limitation, we propose to use a topology optimization algorithm coupled with a state-of-the-art Monte-Carlo transport code. This approach recently proved capable of finding complex optimal configurations of particle propagators with limited human intervention. Main results. In this study, we apply this algorithmic solution to optimize some heavy-water neutron moderators for a specific AB-BNCT treatment unit. The moderators thus generated are compact yet succeed in limiting the exposure of patient's healthy tissues to levels below recommended limits. They present subtle, original geometries inaccessible to standard parametric approaches or human intuition. Significance. This approach could be used to automatically fit the design of a BNCT moderator to the location and shape of the tumor or to the morphology of the patient to be treated, opening a path for more targeted BNCT treatment.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据