Improving small animal cone beam CT resolution by mitigating x-ray focal spot induced blurring via deconvolution

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 67, 期 12, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1361-6560/ac6b7a

关键词

cone beam CT; image resolution; x-ray spot size; deconvolution

资金

  1. National Institutes of Health [R37CA214639, R01CA227289]

向作者/读者索取更多资源

This study proposes a simple and effective method to mitigate the blurring effect caused by the finite size of the x-ray focal spot and improve the resolution of cone beam computerized tomography (CBCT) images. By measuring the intensity distribution of the x-ray focal spot and deconvolving the CBCT images, the spatial resolution of the images is successfully improved, leading to enhanced visualization of fine animal anatomical structures.
Objective. Modern preclinical small animal radiation platforms utilize cone beam computerized tomography (CBCT) for image guidance and experiment planning purposes. The resolution of CBCT images is of particular importance for visualizing fine animal anatomical structures. One major cause of spatial resolution reduction is the finite size of the x-ray focal spot. In this work, we proposed a simple method to measure x-ray focal spot intensity map and a CBCT image domain deblurring model to mitigate the effect of focal spot-induced image blurring. Approach. We measured a projection image of a tungsten ball bearing using the flat panel detector of the CBCT platform. We built a forward blurring model of the projection image and derived the spot intensity map by deconvolving the measured projection image. Based on the measured spot intensity map, we derived a CBCT image domain blurring model for images reconstructed by the filtered backprojection algorithm. Based on this model, we computed image domain blurring kernel and improved the CBCT image resolution by deconvolving the CBCT image. Main results. We successfully measured the x-ray focal spot intensity map. The spot size characterized by full width at half maximum was similar to 0.75 x 0.55 mm(2) at 40 kVp. We computed image domain convolution kernels caused by the x-ray focal spot. A simulation study on noiseless projections was performed to evaluate the spatial resolution improvement exclusively by the focal spot kernel, and the modulation transfer function (MTF) at 50% was increased from 1.40 to 1.65 mm(-1) for in-plane images and 1.05-1.32 mm(-1) for cross-plane images. Experimental studies on a CT insert phantom and a plastinated mouse phantom demonstrated improved spatial resolution after image domain deconvolution, as indicated by visually improved resolution of fine structures. MTF at 50% was improved from 1.00 to 1.12 mm(-1) for in-plane direction and from 0.72 to 0.84 mm(-1) for cross-plane direction. Significance. The proposed method to mitigate blurring caused by finite x-ray spot size and improve CBCT image resolution is simple and effective.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据