Successful percutaneous extraction of malpositioned pacemaker lead in the left ventricle after proper dabigatran treatment

期刊

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
卷 45, 期 9, 页码 1101-1105

出版社

WILEY
DOI: 10.1111/pace.14491

关键词

malpositioned pacemaker lead; percutaneous lead extraction; thromboembolic complication

资金

  1. Eva and Carl-Eric Larsson foundation

向作者/读者索取更多资源

We present a case of transient ischemic attacks after pacemaker implantation due to the malpositioned pacemaker lead in the left ventricle. The misplaced ventricular lead was identified during regular echocardiography and successfully percutaneously extracted after dabigatran treatment. No neurologic events were observed during a 4-year follow-up.
Malpositioned pacemaker lead in the left ventricle (LV) is a rare procedural complication, which causes a special risk of thromboembolic events. Hence, prompt identification and early management of misplaced leads inside the LV is critical. Herein, we present a case of malpositioned pacemaker lead with transient ischemic attacks after the pacemaker implantation. The misplaced ventricular lead was discovered during regular echocardiography. Both leads were extracted percutaneously after dabigatran treatment. To our knowledge, this is the first report of uncomplicated percutaneous extraction of an inadvertently placed LV lead after dabigatran treatment. No neurologic events during a follow-up of 4 years.