4.7 Article

A voltammetric approach to the quantification of tungsten in environmental waters using a solid bismuth microelectrode

期刊

MEASUREMENT
卷 194, 期 -, 页码 -

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.measurement.2022.111089

关键词

Tungsten; Electrochemical measurement; Stripping voltammetry; Miniaturized sensor; Real water sample; Resin

向作者/读者索取更多资源

This paper discusses a novel eco-friendly solid bismuth microelectrode for direct tungsten quantification. The microelectrode used in this study exhibits perfect long-term persistence and insensitivity to the presence of oxygen in the solution. The sensitivity of the sensor was optimized by investigating instrumental and chemical factors. The method shows good linearity and selectivity, and it has been successfully applied to the analysis of various samples.
In this paper, a novel eco-friendly solid bismuth microelectrode for direct tungsten quantification by cathodic stripping voltammetry (CSV) is discussed. The novelty of the proposed sensor results from the fact that the used microelectrode (o = 25 mu m) exhibits perfect long-term persistence. Moreover, on account of spherical diffusion and insensitivity to the presence of oxygen in the solution the analysis could be performed at the point of sampling. The instrumental and chemical factors influencing the sensitivity of the sensor were thoroughly investigated and their most favorable values were selected (pH = 2.88, acetic acid concentration: 0.1 M; activation potential/time:-1.8 V/1 s; accumulation potential/time:-0.5 V/20 s). Under selected conditions, the method exhibits linearity in the range of 1 x 10(-8) to 5 x 10(-6) M, while the detection boundary was calculated to be 1.3 x 10(-9) M. This method has been shown to be highly selective and insensitive to the presence of even 10 mg L-1 of humic substances due to the addition of the adsorptive Amberlite XAD-7 resin to the test solution. In keeping with the outlined procedure, the certified reference materials (SPS-SW1, SPS-WW1, NAAS-5), tap and well water samples were analyzed without any pretreatment and recovery values from 90.4 to 108.5% were obtained.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据