Evaluation of the sample-to-result, random access NeuMoDx platform for viral load testing of Cytomegalovirus and Epstein Barr virus in clinical specimens

期刊

JOURNAL OF CLINICAL VIROLOGY
卷 149, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.jcv.2022.105122

关键词

EBV; CMV; PCR; NeuMoDx; Viral load; Fully automated

类别

向作者/读者索取更多资源

This study evaluated the qualitative and quantitative performance of the NeuMoDx CMV and EBV assays compared to lab developed tests (LDT) and assays on the Abbott m2000 system. The results showed that the NeuMoDx assays are suitable for medium-to-high throughput diagnostic settings, but there are differences in sensitivity and quantification compared to the Abbott m2000 system, especially for EBV.
Background: The detection and follow up of cytomegalovirus (CMV) and Epstein-Barr virus (EBV) viral loads (VL) are crucial in the management of immunocompromised patients. Recently, molecular CE-IVD assays for detection and quantification of CMV and EBV have been launched for use on the random-access and sample-to-result NeuMoDx 96 and 288 platforms (Qiagen). Objective: Evaluating the qualitative and quantitative performance of the NeuMoDx CMV and EBV assays in clinical specimens compared to a lab developed tests (LDT) and the CE-IVD assays on the Abbott m2000 system. Method: Both a prospective and a retrospective panel, compiled of non-detectable (ND), non-quantifiable (NQ) and quantifiable VLs in plasma samples have been evaluated for both CMV and EBV: NeuMoDx versus LDT and NeuMoDx versus Abbott m2000. Quantitative agreement was determined for samples with a quantifiable VL on both systems. Results: Qualitative and quantitative agreement between the NeuMoDx and LUMC's LDT CMV assays was 88%. Qualitative agreement between the NeuMoDx and Abbott m2000 CMV assays was 92% and quantitative agreement was 87%. Qualitative and quantitative agreement between the NeuMoDx and the LDT EBV assays was 87%. Qualitative agreement between the NeuMoDx and Abbott m2000 EBV assays was 91% and quantitative agreement was 0%. Conclusion: These data show that the NeuMoDx assays are suitable for both detection and quantification of CMV and EBV in a medium-to high throughput diagnostic setting, but that differences in sensitivity and quantification (for EBV, NeuMoDx versus Abbott m2000) warrant an extensive transition period when using the respective assays for following VL in patient samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据