4.6 Article

Development of a validated method for rapid quantification of up to sixteen cannabinoids using ultra-high-performance liquid chromatography diode-array detector with optional electrospray ionization time-of-flight mass spectrometry detection

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1670, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2022.462953

关键词

Cannabis; Cannabinoids; Electrospray ionization time-of-flight mass; spectrometry; Hemp concentrates; Ultra-high-performance liquid; chromatography diode-array detector

资金

  1. National Institute of Justice [2020-DQ-BX-0021]
  2. National Science Foundation [CHE-1827209]
  3. Western Illinois University

向作者/读者索取更多资源

Due to the legalization of medical and recreational Cannabis in many countries, there is an increasing demand for accurate quantification of cannabinoids. This study developed a rapid quantification method for multiple cannabinoids and achieved good separation in hemp concentrate analysis. The results showed the absence of acidic cannabinoids but the presence of neutral cannabinoids, with some samples containing high levels of Delta(9)-THC. Additionally, the issue of recovery experiments was addressed by using abnormal CBD as a spike and untargeted cannabinoids were discovered.
Due to the recent legalization of medical and recreational Cannabis in many countries in the world, there has been an increasing demand for accurate quantification of a growing number of cannabinoids. To meet this challenge, a method for rapid quantification of up to sixteen cannabinoids using ultra-high-performance liquid chromatography diode-array detector (UHPLC-DAD) has been developed, validated and used in the analysis of hemp concentrates. While published LC-UV methods were usually for twelve or less cannabinoids and might not achieve baseline separation of some critical pairs of cannabinoids, e.g., CBG/CBD (cannabigerol/cannabidiol) and Delta(9)-THC/Delta(8)-THC (tetrahydrocannabinol), in this study a systematic separation optimization led to a resolution of 1.7 for both pairs. The linear calibration range of all cannabinoids were between 0.02 to 25 mu g/mL in methanol, leading to the quantification of 0.1 to 125% (w/w) individual cannabinoids in hemp concentrates after they were mixed with methanol at 20 mu g/mL and analyzed after ultrasonication, centrifugation and filtration. The analytical results showed that none of the nine analyzed samples contained any of the seven acidic cannabinoids, but all the nine neutral cannabinoids were detected, with average content ranging from 0.10 to 93.23% (w/w) and RSD (relative standard deviation) values from 0.3 to 11.2% in triplicates. Particularly, two hemp concentrates, i.e., delta 8 hemp distillate and delta 8 hemp shatter contained 9.35 and 11.33% (w/w) Delta(9)-THC, respectively. While published recovery experiments were limited by the unavailability of cannabinoid-free matrix and the high cost of cannabinoid standards, this problem was solved by spiking abnormal CBD, a cannabinoid not naturally present in Cannabis plants and commercially available with a reasonable price, into the samples. The obtained average recovery ranged from 94.8 to 103.6% with RSD values from 1.5 to 9.0% in triplicates for the nine analyzed samples. Electrospray ionization time-of-flight mass spectrometry (ESI/TOFMS) confirmed the good specificity of the UHPLC-DAD method, i.e., without any false positive identification of individual cannabinoids, and discovered six untargeted cannabinoids that were structural isomers of Delta(9)-THC in the nine hemp concentrate samples. (c) 2022 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。