4.2 Article

Development and Validation of a Quantitative Method for Multiple Allergen Detection in Food Using Concatemer-Based Isotope Dilution Mass Spectrometry

期刊

JOURNAL OF AOAC INTERNATIONAL
卷 105, 期 6, 页码 1585-1595

出版社

OXFORD UNIV PRESS INC
DOI: 10.1093/jaoacint/qsac053

关键词

-

资金

  1. Belgian Federal Public Service of Health, Food Chain Safety and Environment [RT 15/10 ALLERSENS]

向作者/读者索取更多资源

This study has successfully developed a robust method for detecting and quantifying multiple allergens in food, ensuring the safety of allergic consumers. The combination of specific incurred test materials, standard addition strategy, and stable isotope-labeled concatemer has provided the method with sufficient sensitivity and reliability.
Background Accurate food labeling is essential to protect allergic consumers. However, allergen contaminations may occur during the whole food production process. Reliable, sensitive, and robust methods for detecting multiple allergens in food are needed. Objective This work aims to develop and validate an LC coupled to tandem mass spectrometry (MS/MS) method for the detection and quantification of hazelnuts, peanuts, milk, and eggs in processed food products. Methods In-house-produced incurred test materials, cookies and chocolates, were used for the method development and validation. The quantification was based on the standard addition strategy using qualified reference materials as allergen protein standards and an innovative stable isotope-labeled concatemer as an internal standard. Results A method targeting 19 allergen-specific peptides was developed and validated in two laboratories, which strengthens its robustness. The AOAC INTERNATIONAL performance requirements for repeatability, intermediate precision, reproducibility, and recovery were reached for at least one peptide per allergen across both matrixes, and quantification limits complied with the action levels of the Food Industry Guide to the Voluntary Incidental Trace Allergen Labelling (VITAL (R)) Program Version 3.0. Conclusion The combination of incurred test materials, standard addition strategy, and stable isotope-labeled concatemer as an internal standard allowed us to develop and validate a robust method for detecting and quantifying multiple allergens in food with sufficient sensitivity to protect allergic consumers. Highlights: The combination of characterized incurred test material, calibration with certified reference material, a single stable isotope labelled concatemer and cross-lab validation result in the required standardization and harmonization in food allergen detection according to the stakeholders' group to assess the robustness of our method.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据