A needle deflection model with operating condition optimization for corrective force-based needle guidance during transrectal prostate brachytherapy

出版社

WILEY
DOI: 10.1002/rcs.2388

关键词

correction force; deflection estimation; needle insertion; optimization experiment; puncture accuracy; transrectal brachytherapy

类别

资金

  1. Key Projects of Natural Science Foundation of Heilongjiang Province of China [ZD2018013]
  2. National Natural Science Foundation of China [51675142]
  3. Reserve Leader Funding Project of Leading Talent Echelon of Heilongjiang Province of China [2501050628]

向作者/读者索取更多资源

This study investigates the use of an auxiliary needle insertion guide based on corrective forces for transrectal prostate brachytherapy. The guide effectively reduces needle deflection and its efficacy is experimentally verified. Optimal puncture parameters are determined through optimization, resulting in improved puncture accuracy.
Background Transrectal prostate brachytherapy (BT) involves the permanent implantation of radioactive seeds through a needle, which must be inserted along a straight path to satisfy dose distribution requirements. However, this process is complicated by bevelled needle tips that can cause deflection during penetration. In clinical practice, physicians typically rotate the bevelled tip intermittently or apply manual correction forces near the insertion point, to reduce needle deflection. While assisted rotational insertion robots have made substantial progress in the past 20 years, tissue sticking can be caused by rotation of the bevelled tip and there are currently few studies on the use of corrective forces. As such, an auxiliary needle insertion guide for transrectal prostate BT, based on corrective forces, is investigated in this study for the first time. Methods The proposed BT guide is designed to reduce needle deflection and was experimentally verified by in vitro experiments. An energy-based deflection model was developed to predict needle motion as corrective forces were applied during insertion. An experimental platform was constructed to perform corrective force-assisted punctures, using the magnitude of the corrective force (A), the position of corrective force application (B) and the puncture speed (C) as test factors, with needle deflection as a test indicator. Design-Expert 8.0.5b software was used for simulation, and a high-definition camera acquired pictures of the needle tip as it pierced the tissue. A regression equation was also established for the test factors and test indicators, using Design-Expert software. Optimal parameter combinations (A, B and C) were determined through optimization. Results Calculated needle deflection values were then compared with the measured position of the needle insertion point, producing an average error of 0.39 +/- 0.28 mm. Deflection was as low as 0.8 mm using optimal puncture parameters Conclusions A needle-tissue interaction model, considering tissue nonlinearity, which experimental results demonstrated to be highly accurate. Optimal puncture parameters effectively improved puncture accuracy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据