4.7 Article

Patient-specific multi-scale design optimization of transcatheter aortic valve stents

期刊

出版社

ELSEVIER IRELAND LTD
DOI: 10.1016/j.cmpb.2022.106912

关键词

Transcatheter aortic valve; Stent; Finite element method; Multi-objective optimization; Gaussian process regression models

资金

  1. Italian Minstry of Education, University and Research [1613 FISR2019_03221]

向作者/读者索取更多资源

The design and optimization of transcatheter heart valve replacement devices are of great importance for improving performance quality and reducing the risk of malfunction. In this study, optimal models with larger distal diameters were found to perform better in the selected objective functions.
Background and Objective:Transcatheter aortic valve implantation (TAVI) has become the standard treatment for a wide range of patients with aortic stenosis. Although some of the TAVI post-operative complications are addressed in newer designs, other complications and lack of long-term and durability data on the performance of these prostheses are limiting this procedure from becoming the standard for heart valve replacements. The design optimization of these devices with the finite element and optimization techniques can help increase their performance quality and reduce the risk of malfunctioning. Most performance metrics of these prostheses are morphology-dependent, and the design and the selection of the device before implantation should be planned for each individual patient. Methods:In this study, a patient-specific aortic root geometry was utilized for the crimping and implantation simulation of 50 stent samples. The results of simulations were then evaluated and used for developing regression models. The strut width and thickness, the number of cells and patterns, the size of stent cells, and the diameter profile of the stent were optimized with two sets of optimization processes. The objective functions included the maximum crimping strain, radial strength, anchorage area, and the eccentricity of the stent. Results:The optimization process was successful in finding optimal models with up to 40% decrease in the maximum crimping strain, 261% increase in the radial strength, 67% reduction in the eccentricity, and about an eightfold increase in the anchorage area compared to the reference device. Conclusions: The stents with larger distal diameters perform better in the selected objective functions. They provide better anchorage in the aortic root resulting in a smaller gap between the device and the surrounding tissue and smaller contact pressure. This framework can be used in designing patient-specific stents and improving the performance of these devices and the outcome of the implantation process. (c) 2022 Published by Elsevier B.V.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据