4.6 Article

Transcarotid Transcatheter Aortic Valve Replacement Feasibility and Safety

期刊

JACC-CARDIOVASCULAR INTERVENTIONS
卷 9, 期 5, 页码 472-480

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcin.2015.11.045

关键词

aortic stenosis; carotid vascular access; transcatheter aortic valve replacement

向作者/读者索取更多资源

OBJECTIVES The purpose of this study was to assess the feasibility and safety of transcarotid transcatheter aortic valve replacement (TAVR). BACKGROUND Many candidates for TAVR have challenging vascular anatomy that precludes transfemoral access. Transcarotid arterial access may be an option for such patients. METHODS The French Transcarotid TAVR Registry is a voluntary database that prospectively collected patient demographics, procedural characteristics, and clinical outcomes among patients undergoing transcarotid TAVR. Outcomes are reported according to the updated Valve Academic Research Consortium criteria. RESULTS Among 96 patients undergoing transcarotid TAVR at 3 French sites (2009 to 2013), the mean age and Society of Thoracic Surgeons predicted risk of mortality were 79.4 + 9.2 years and 7.1 + 4.1%, respectively. Successful carotid artery access was achieved in all patients. The Medtronic CoreValve (Medtronic, Inc., Minneapolis, Minnesota) (n = 89; 92.7%) and Edwards SAPIEN valves (Edwards Lifesciences, Irvine, California) (n = 7; 7.3%) were used. Procedural complications included: valve embolization (3.1%), requirement for a second valve (3.1%), and tamponade (4.2%). There were no major bleeds or major vascular complications related to the access site. There were 3 (3.1%) procedural deaths and 6 (6.3%) deaths at 30 days. The 1-year mortality rate was 16.7%. There were 3 (3.1%) cases of Valve Academic Research Consortium-defined in-hospital stroke (n = 0) or transient ischemic attack (TIA) (n = 3). None of these patients achieved the criteria for stroke and none manifested new ischemic lesions on cerebral computed tomography or magnetic resonance imaging. At 30 days, a further 3 TIAs were observed, giving an overall stroke/TIA rate of 6.3%. CONCLUSIONS Transcarotid vascular access for TAVR is feasible and is associated with encouraging short-and medium-term clinical outcomes. Prospective studies are required to ascertain if transcarotid TAVR yields equivalent results to other nonfemoral vascular access routes. (C) 2016 by the American College of Cardiology Foundation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据