4.6 Article

subclass analysis using dual detection gel-permeation high-performance liquid chromatography system

期刊

BIOSCIENCE REPORTS
卷 42, 期 6, 页码 -

出版社

PORTLAND PRESS LTD
DOI: 10.1042/BSR20220291

关键词

-

资金

  1. Japan Society for the Promotion of Science [21K15641, 19K07932]
  2. Grants-in-Aid for Scientific Research [21K15641, 19K07932] Funding Source: KAKEN

向作者/读者索取更多资源

The LipoSEARCH System is an innovative method for analyzing lipoprotein subclasses, and a developed internal standard using a pigment has been shown to guarantee simultaneous detection and improve accuracy.
The LipoSEARCH?? System is an innovative lipoprotein class analysis method based on gel-permeation high-performance liquid chromatography (HPLC). This system uses a gel permeation column to separate the major lipoprotein subclasses (chylomicron, very low-density lipoprotein, low-density lipoprotein, and high-density lipoprotein) in serum according to particle size and splits them into two pathways to measure total cholesterol (TC; esterified + unesterified cholesterol) and triglyceride (TG) concentrations simultaneously to obtain chromatograms for each. These chromatograms were analyzed based on the results of the calibration serum by fitting Gaussian curves to profile the 20 lipoprotein subclasses defined in detail. An important assumption of this HPLC system is its simultaneous detection of two pathways to guarantee the accuracy of each analysis. Therefore, in the present study, we investigated the development of an internal standard that can guarantee the simultaneous detection of this system by adding a pigment to the serum. We focused on quinone pigments with absorption at 550 nm, which is the wavelength used for the enzymatic assay of TC and TG concentrations in the system. As a result, we succeeded in producing overlapping pigment peaks that appeared after the analytical chromatograms in two pathways. It is also suggested that the pigment solution as an internal standard is stable in freezing storage and has little effect on the analysis. The developed internal standard is expected to contribute to the accuracy assurance of lipoprotein analysis by this dual-detection HPLC system.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据