Dual-color quantum dot-loaded nanoparticles based lateral flow biosensor for the simultaneous detection of gastric cancer markers in a single test line

期刊

ANALYTICA CHIMICA ACTA
卷 1218, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.aca.2022.339998

关键词

Simultaneous quantitative detection; Dual-color quantum dot-loaded particles; Lateral flow immunoassay; Pepsinogen ratio

资金

  1. National Natural Science Foundation of China [81872416]
  2. China Postdoctoral Science Founda-tion [2019M662989]

向作者/读者索取更多资源

The dual-color quantum dots-loaded nanoparticles based lateral flow biosensor offers high accuracy by eliminating heterogeneity caused by different membrane locations, showing success in detecting pepsinogen with ultra-low limits of detection. The capability and reliability of the multiplex biosensor are validated, demonstrating negligible crosstalk and cross-reaction during simultaneous detection.
A dual-color quantum dots-loaded nanoparticles (QPs) based lateral flow biosensor with single test line has been developed. Red and green emitted QPs were conjugated with antibodies and served as detecting probes in assays respectively, while the mixture of various antibodies were immobilized on nitrocellulose membranes as one detection line. Benefit from eliminating the heterogeneity caused by different position on the membrane, current biosensor achieved higher accuracy comparing with prevalent multi-lines or multi-strips lateral flow systems, which is of great significance for analyzing ratio-related diagnostics. The capability and reliability of the multiplex biosensor are also demonstrated by utilizing pepsinogen I (PG I) and pepsinogen II (PG II) as the model analytes. Under the optimal conditions, quantitative detection was achieved with ultra-low limits of detection at 6.9 pM (0.29 ng mL-1, PG I) and 15.7 pM (0.66 ng mL-1, PG II) respectively. The spectra crosstalk was negligible and no apparent cross-reaction was found in simultaneous detection. Furthermore, a good linear correlation of the QPs based lateral flow biosensor and commercial time-resolved fluoroimmunoassay was obtained in the detection of clinical samples, indicating the high reliability of the proposed biosensor.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据