This study aimed to develop a simple and rapid analytical method using RP-HPLC for the estimation of Metronidazole, Ciprofloxacin and Doxycycline in bulk and pharmaceutical dosage form. The method was successfully validated and applied for the analysis of these drugs, showing good accuracy and precision.
Aim: To design simple, rapid, new analytical method for estimation of Metronidazole, Ciprofloxacin Doxycycline by using RP-HPLC in bulk and pharmaceutical dosage form. Study Design: Estimation of Metronidazole, Ciprofloxacin Doxycycline by using RP-HPLC in bulk and pharmaceutical dosage form was planned and executed. Place and Duration of Study: Chalapathi Drug Testing Laboratory, Chalapathi Institute Of Pharmaceutical Sciences, Lam, Guntur-522034, Andhra Pradesh, India during the period of November 2019 to February 2020. Methodology: Metronidazole is an antibiotic and antiprotozoal medication. It is used either alone or with other antibiotics to treat pelvic inflammatory disease, endocarditis and bacterial vaginosis. Ciprofloxacin is an antibiotic used to treat a number of bacterial infections. Doxycycline is a tetracycline antibiotic that fights bacteria in the body. The study was carried out on SHIMADZU Prominance-i, LC-2030C system equipped with Shim-pack Gist (250 x 4.6 mm, 5 mu m) column and mobile phase was optimized by using mixture of methanol and 0.25mM potassium phosphate buffer in the ration of 60:40 v/v at a flow rate of 0.8 ml/min. The wavelength was set as 282nm at ambient temperature by injecting 20 mu l of solution and the run time was fixed for 10 min. Results: Linearity plot was constructed for concentration range of 5-15 mu g/ml for Metronidazole,515 mu g/ml for Ciprofloxacin and 1-8 mu g/ml for Doxycycline standard solutions. It shows best regression coefficient and y/s values. The accuracy of the proposed method was determined by performing recovery studies and was found between 98-102%. The repeatability testing for both sample and standard solutions the %RSD was found as <2.0% which is within the acceptable limits showing that the method is precise as well. The LOD and LOQ were found to be 0.33 and 0.99 mu g/ml for Metronidazole, 0.33 and 0.99 mu g/ml for Ciprofloxacin, 0.08 and 0.25 mu g/ml for Doxycycline respectively. The proposed method was successfully applied for the pharmaceutical dosage form of Metronidazole, Ciprofloxacin Doxycycline and it was as economic, eco-friendly with less retention time around 10.0 min. Conclusion: The proposed method was validated in terms of linearity, range, Accuracy, precision, Specificity, Robustness. Method was successfully applied to the estimation of Metronidazole, Ciprofloxacin Doxycycline in combined marketed pharmaceutical dosage form.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据