4.2 Article Proceedings Paper

Verification and quality control of routine hematology analyzers

期刊

出版社

WILEY
DOI: 10.1111/ijlh.12503

关键词

Hematology Analyzer; Verification; Complete Blood Count

资金

  1. Abbott Diagnostics (Santa Clara, CA, USA)

向作者/读者索取更多资源

Verification of hematology analyzers (automated blood cell counters) is mandatory before new hematology analyzers may be used in routine clinical care. The verification process consists of several items which comprise among others: precision, accuracy, comparability, carryover, background and linearity throughout the expected range of results. Yet, which standard should be met or which verification limit be used is at the discretion of the laboratory specialist. This paper offers practical guidance on verification and quality control of automated hematology analyzers and provides an expert opinion on the performance standard that should be met by the contemporary generation of hematology analyzers. Therefore (i) the state-of-the-art performance of hematology analyzers for complete blood count parameters is summarized, (ii) considerations, challenges, and pitfalls concerning the development of a verification plan are discussed, (iii) guidance is given regarding the establishment of reference intervals, and (iv) different methods on quality control of hematology analyzers are reviewed.