Validation and Concordance Analysis of a New Lateral Flow Assay for Detection of Histoplasma Antigen in Urine

期刊

JOURNAL OF FUNGI
卷 7, 期 10, 页码 -

出版社

MDPI
DOI: 10.3390/jof7100799

关键词

histoplasmosis; Histoplasma; antigen; HIV; AIDS

资金

  1. Global Disease Detection Program
  2. CDC, Atlanta, GA, USA
  3. Corporacion para Investigaciones Biologicas (CIB, Medellin, Colombia)
  4. Colciencias, Bogota, Colombia via the Program for Young Investigators and Innovators
  5. Fondo de Investigaciones de la Universidad del Rosario (FIUR), Bogota, Colombia
  6. Oak Ridge Institute for Science and Education (ORISE, Oak Ridge, TN, USA)

向作者/读者索取更多资源

The study validated the MiraVista(R) Diagnostics urine antigen lateral flow assay for detecting Histoplasma capsulatum antigen and compared it against quantitative enzyme immunoassays. The MVD LFA showed high sensitivity and specificity, with a quick turnaround time for results and ease of use.
HIHistoplasmosis is a major cause of mortality in people living with HIV (PLHIV). Rapid methods to diagnose Histoplasma capsulatum disease could dramatically decrease the time to initiate treatment, resulting in reduced mortality. The aim of this study was to validate a MiraVista(R) Diagnostics (MVD) Histoplasma urine antigen lateral flow assay (MVD LFA) for the detection of H. capsulatum antigen (Ag) in urine and compare this LFA against the MVista(R) Histoplasma Ag quantitative enzyme immunoassays (MVD EIA). We assessed the MVD LFA using a standardized reference panel of urine specimens from Colombia. We tested 100 urine specimens, 26 from PLHIV diagnosed with histoplasmosis, 42 from PLHIV with other infectious diseases, and 32 from non-HIV infected persons without histoplasmosis. Sensitivity and specificity of the MVD LFA was 96%, compared with 96% sensitivity and 77% specificity of the MVD EIA. Concordance analysis between MVD LFA and the MVD EIA displayed an 84% agreement, and a Kappa of 0.656. The MVD LFA evaluated in this study has several advantages, including a turnaround time for results of approximately 40 min, no need for complex laboratory infrastructure or highly trained laboratory personnel, use of urine specimens, and ease of performing.