Performance evaluation of a PET insert for preclinical MRI in stand-alone PET and simultaneous PET-MRI modes

期刊

EJNMMI PHYSICS
卷 8, 期 1, 页码 -

出版社

SPRINGER
DOI: 10.1186/s40658-021-00415-1

关键词

Preclinical PET; PET insert; PET; MRI; Simultaneous PET; MRI; NEMA

资金

  1. NSW Government Research Attraction and Acceleration Program
  2. Australian National Imaging Facility through the National Collaborative Research Infrastructure Strategy (NCRIS)

向作者/读者索取更多资源

This study evaluated the performance of a preclinical PET insert in three configurations: stand-alone, inside a 3T MRI bore, and during simultaneous PET/MRI scans. Results showed that the performance of the PET insert and MRI system was not significantly affected by the scanning mode.
Background This study aimed to evaluate the performance of a preclinical PET insert in three configurations: as a stand-alone unit outside the MRI bore, inside the bore of a cryogen-free 3T MRI and, finally, while performing simultaneous PET/MRI studies. Methods The PET insert consists of two rings of six detectors, each detector comprising 8 x 12 SiPMs reading out dual offset layers of pixelated LYSO crystals with a 1.4-mm pitch. The inner diameter is 60 mm, transaxial field of view (FoV) 40 mm and axial FoV 98 mm. Evaluation was based on NEMA NU 4-2008 guidelines with appropriate modifications. Spatial resolution and sensitivity were measured inside and outside the MR bore. Image quality, count rate and quantitative performance were measured in all three configurations. The effect of temperature stability on PET sensitivity during fast spin echo sequences was also evaluated. B-0 field homogeneity and T1 and T2 relaxation times were measured using a water-filled phantom, with and without simultaneous PET operation. Finally, PET and MRI scans of a mouse injected with 10 MBq [F-18]NaF and a mouse injected with 16 MBq [F-18]FDG were performed in sequential and simultaneous modes. Results Peak absolute sensitivity was 10.15% with an energy window of 250-750 keV. Absolute sensitivity values outside and inside the MR bore with MR idle agreed to within 0.1%. Outside the MR bore, spatial resolution was 1.21/1.59 mm FWHM (radial/tangential) 5 mm from the centre of the FoV which compared well with 1.19/1.26 mm FWHM inside the MR bore. There were no substantial differences between all three scan configurations in terms of peak NEC rate (175 kcps at 17 MBq), scatter or random fractions. Uniformity and recovery coefficients were also consistent between scanning modes. B-0 field homogeneity and T1 and T2 relaxation times were unaltered by the presence of the PET insert. No significant differences were observed between sequential and simultaneous scans of the animals. Conclusions We conclude that the performance of the PET insert and MRI system is not significantly affected by the scanning mode.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据