4.6 Article

Five-Years Angiographic Follow-Up of Wide-Neck Intracranial Aneurysms Treated With LEO Plus Stent

期刊

FRONTIERS IN NEUROLOGY
卷 12, 期 -, 页码 -

出版社

FRONTIERS MEDIA SA
DOI: 10.3389/fneur.2021.744962

关键词

aneurysm; stent; hemorrhage; embolization; LEO

向作者/读者索取更多资源

This study evaluated the long-term outcome of treating wide-neck intracranial aneurysms with LEO Plus stents, showing high rates of adequate and stable occlusion over time, with low total morbidity and mortality at 5 years post-procedure.
Background: This study aimed to evaluate the angiographic and clinical outcome, with an emphasis on long-term follow-up, of the LEO Plus stent for wide-neck intracranial aneurysms treated in a single center.Methods: We retrospectively examined a prospectively maintained database of patients treated with LEO Plus devices between January 2004 and December 2016. Data regarding patient demographics, aneurysm characteristics, and technical procedures were analyzed. Angiographic and clinical findings were recorded during the procedure and followed up over a period of at least 5 years.Results: We identified 101 patients with 116 aneurysms. In 16 patients, the stent could not safely be placed. Thus, a total of 97 LEO Plus devices were implanted in 97 aneurysms of 85 patients. Adverse events (acute and delayed) were observed in 21.6% of cases (17/85), and most were resolved (70.6%; 12/17). Moreover, 5 years after the procedure, total morbidity and mortality were 2.3% (2/85) and 3.5% (3/85), respectively. Long-term imaging follow-up showed complete occlusions, neck remnants, and residual aneurysms in 73.1% (57/78), 14.1% (11/78), and 12.8% (10/78) of cases, respectively.Conclusions: Long-term results of treatment of brain aneurysms with LEO stent show high rates of adequate and stable occlusion over time, with acceptable morbidity and mortality.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据