4.6 Article

Biomechanical Analysis of Staples for Epiphysiodesis

期刊

APPLIED SCIENCES-BASEL
卷 12, 期 2, 页码 -

出版社

MDPI
DOI: 10.3390/app12020614

关键词

biomechanics; orthopedics; Blount staple; FEA; experiment; epiphysiodesis

资金

  1. Motol University Hospital, Prague, Czech Republic [SP2021/66, FV40306, CZ.02.1.01/0.0/0.0/17_049/0008441, CZ.02.1.01/0.0/17_049/0008407, 00064203]

向作者/读者索取更多资源

This article analyzes the performance of Blount bone staples for bone growth arrest and confirms their safety and usability through finite element analysis and experimentation.
Limb asymmetry can, and often does, cause various health problems. Blount bone staples (clips) are used to correct such uneven growth. This article analyzes the performance of a biomechanical staple during bone (tibia) growth arrest. The staples considered in this study were made of 1.4441 stainless steel, the model of tibia consisted of two materials representing corticalis and spongiosis. Hooke's law was used for modeling materials' behaviors for finite element analysis (FEA). The maxima of stress and total staple displacement were evaluated using the finite element method and verification of the results, along with the determination of the maximum loading (growing) force that the staples are capable of withstanding, was performed experimentally. The presented method can be used to determine the safety and usability of staples for bone growth arrest. According to our results, the design of Blount staples considered in this paper is safe and suitable for orthopedic treatment.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。