In this study, the use of iCT was found to significantly reduce radiation exposure to operating room staff, but increase the radiation dose to the patient and prolong operative time in posterior intervertebral lumbar fusion procedures compared to C-arm fluoroscopy.
Navigation systems used during minimally invasive spine procedures have evolved from uniplanar, two-dimensional C-arm fluoroscopy to multiplanar, 3D intraoperative computed tomography (iCT). In this study, the radiation exposure to the patient and operating room staff in posterior intervertebral lumbar fusion procedures is compared between iCT and C-arm fluoroscopy. The effective dose of the surgeon, operating nurse, and anesthesiologist was measured during surgery with personal dosimeters, and the effective dose of the patient was measured with Gafchromic(TM) films. The time efficiency of the procedure was evaluated by recording the duration of pedicle screw fixation and the duration of the total surgery time. A total of 75 patients participated in the study; 30 patients had surgery guided by iCT and 45 by C-arm fluoroscopy. The radiation dose of the surgeon, the operating nurse, and the anesthesiologist was thirteen fold lower with surgeries assisted by iCT compared to C-arm fluoroscopy. In contrast, the effective dose of the patient significantly increased with iCT. Using iCT, radiation exposure of the operating room staff can be significantly reduced. iCT increases the effective dose of the patient and prolongs the operative time.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据