A Cross-Machine Comparison of Shear-Wave Speed Measurements Using 2D Shear-Wave Elastography in the Normal Female Breast

期刊

APPLIED SCIENCES-BASEL
卷 11, 期 20, 页码 -

出版社

MDPI
DOI: 10.3390/app11209391

关键词

shear-wave elastography; shear-wave speed; radiation-induced breast toxicity; tissue stiffness; cross-machine comparison

资金

  1. NHS
  2. Cancer Research UK under Programme [C20892/A23557]
  3. NIHR Cambridge Biomedical Research Centre [BRC-1215-20014]

向作者/读者索取更多资源

This study compared two different shear-wave elastography machines and found relatively large differences in measurements in vivo, which may impact the pooling of data across machines in clinical studies of the breast.
Quantitative measures of radiation-induced breast stiffness are required to support clinical studies of novel breast radiotherapy regimens and exploration of personalised therapy, however, variation between shear-wave elastography (SWE) machines may limit the usefulness of shear-wave speed (c(s)) for this purpose. Mean c(s) measured in four healthy volunteers' breasts and a phantom using 2D-SWE machines Acuson S2000 (Siemens Medical Solutions) and Aixplorer (Supersonic Imagine) were compared. Shear-wave speed was measured in the skin region, subcutaneous adipose tissue and parenchyma. c(s) estimates were on average 2.3% greater when using the Aixplorer compared to S2000 in vitro. In vivo, c(s) estimates were on average 43.7%, 36.3% and 49.9% significantly greater (p << 0.01) when using the Aixplorer compared to S2000, for skin region, subcutaneous adipose tissue and parenchyma, respectively. In conclusion, despite relatively small differences between machines observed in vitro, large differences in absolute measures of shear wave speed measured were observed in vivo, which may prevent pooling of cross-machine data in clinical studies of the breast.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据