ULTRASOUND IN MEDICINE AND BIOLOGY
卷 48, 期 2, 页码 346-357出版社
ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2021.10.004
关键词
Pressure difference estimation; Ultrasound; Vector flow imaging; Fluid-filled pressure catheter; Computational fluid dynamics; Fluid-structure interaction; Carotid bifurcation
This study compared local pressure differences estimated using vector flow imaging and direct catheterization with simulated pressure fields. The results showed a strong correlation between VFI and simulated peak pressure differences, while the catheter method exhibited no correlation. VFI pressure difference estimation was found to be feasible in phantoms and in vivo.
Local pressure differences estimated using vector flow imaging (VFI) and direct catheterization in seven carotid bifurcation phantoms were compared with simulated pressure fields. VFI correlated strongly with simulated peak pressure differences (r = 0.99, p < 0.00001), with an average overestimation of 12.3 Pa (28.6%). The range between the lowest and highest pressure difference of VFI underestimated simulations by 4.6 Pa (8.06%; r = 0.99, p < 0.0001). The catheter method exhibited no correlation (r = -0.09, p = 0.85). Ten repeated measurements on one phantom revealed a small standard deviation (SD) for VFI (SD = 8.4%, mean estimated SD = 11.5%), but not for the catheter method (SD = 785.6%). An in vivo peak systolic pressure difference of 97.9 Pa (estimated SD = 30%) was measured using VFI in one healthy individual. This study indicates that VFI pressure difference estimation is feasible in phantoms and in vivo and realistic estimates of the SD can be attained from the data. (C) 2021 World Federation for Ultrasound in Medicine & Biology. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据