Initial results of a mouse brain PET insert with a staggered 3-layer DOI detector

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 66, 期 21, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1361-6560/ac311c

关键词

positron emission tomography; high-resolution; depth-of-interaction

资金

  1. Japan Society for the Promotion of Science (JSPS) KAKENHI Grants [20H05667, 21K19936]
  2. Grants-in-Aid for Scientific Research [20H05667, 21K19936] Funding Source: KAKEN

向作者/读者索取更多资源

A mouse brain PET prototype was successfully developed with a staggered 3-layer DOI detector, achieving clear identification of brain structures with high contrast. The use of DOI detector improved resolution and reduced inter-crystal scatter, resulting in good imaging performance.
Objective. Small animal positron emission tomography (PET) requires a submillimeter resolution for better quantification of radiopharmaceuticals. On the other hand, depth-of-interaction (DOI) information is essential to preserve the spatial resolution while maintaining the sensitivity. Recently, we developed a staggered 3-layer DOI detector with 1 mm crystal pitch and 15 mm total crystal thickness, but we did not demonstrate the imaging performance of the DOI detector with full ring geometry. In this study we present initial imaging results obtained for a mouse brain PET prototype developed with the staggered 3-layer DOI detector. Approach. The prototype had 53 mm inner diameter and 11 mm axial field-of-view. The PET scanner consisted of 16 DOI detectors each of which had a staggered 3-layer LYSO crystal array (4/4/7 mm) coupled to a 4 x 4 silicon photomultiplier array. The physical performance was evaluated in terms of the NEMA NU 4 2008 protocol. Main Results. The measured spatial resolutions at the center and 15 mm radial offset were 0.67 mm and 1.56 mm for filtered-back-projection, respectively. The peak absolute sensitivity of 0.74% was obtained with an energy window of 400-600 keV. The resolution phantom imaging results show the clear identification of a submillimetric rod pattern with the ordered-subset expectation maximization algorithm. The inter-crystal scatter rejection using a narrow energy window could enhance the resolvability of a 0.75 mm rod significantly. Significance. In an animal imaging experiment, the detailed mouse brain structures such as cortex and thalamus were clearly identified with high contrast. In conclusion, we successfully developed the mouse brain PET insert prototype with a staggered 3-layer DOI detector.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据