4.4 Article

Gold coated contact lens-type ocular in vivo dosimeter (CLOD) for monitoring of low dose in computed tomography: A Monte Carlo study

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2021.10.016

关键词

Contact lens dosimeter; Lens dose; Dose enhancement; Monte Carlo simulation; Gold nanoparticle

资金

  1. National Research Foundation of Korea (NRF) , Republic of Korea - Korea government (MSIT) [2019K1A3A1A80113183]
  2. Radiation Technology R&D program through the National Research Foundation of Korea - Korea government (MIST) [2019M2A2B4095126]
  3. National Research Foundation of Korea [2019M2A2B4095126, 2019K1A3A1A80113183] Funding Source: Korea Institute of Science & Technology Information (KISTI), National Science & Technology Information Service (NTIS)

向作者/读者索取更多资源

The study demonstrated a sensitivity enhancement of gold-coated CLODs for low-dose measurements in CT imaging. The dose enhancement factor decreased as the thickness of the active layer increased. Gold-coated CLODs showed potential for evaluating very low doses delivered to patients during CT imaging.
Purpose: This study reports a sensitivity enhancement of gold-coated contact lens-type ocular in vivo dosimeters (CLODs) for low-dose measurements in computed tomography (CT). Methods: Monte Carlo (MC) simulations were conducted to evaluate the dose enhancement from the gold (Au) layers on the CLODs. The human eye and CLODs were modeled, and the X-ray tube voltages were defined as 80, 120, and 140 kVp. The thickness of the Au layer attached to a CLOD ranged from 100 nm to 10 mu m. The thickness of the active layer ranged from 20 to 140 mu m. The dose ratio between the active layer of the Au-coated CLOD and a CLOD without a layer, i.e., the dose enhancement factor (DEF), was calculated. Results: The DEFs of the first 20-mu m thick active layer of the 5-mu m thick Au-coated CLOD were 18.4, 19.7, 20.2 at 80, 120, and 140 kVp, respectively. The DEFs decreased as the thickness of the active layer increased. The DEFs of 100-nm to 5-mu m thick Au layers increased from 1.7 to 5.4 for 120-kVp X-ray tube voltage when the thickness of the active layer was 140 mu m. Conclusions: The MC results presented a higher sensitivity of Au-coated CLODs (similar to 20-times higher than that of CLODs without a gold layer). Au-coated CLODs can be applied to an evaluation of very low doses (a few cGy) delivered to patients during CT imaging.