4.3 Article

Feasibility of a Novel ECG Electrode Placement Method in Newborn Infants

期刊

NEONATOLOGY
卷 119, 期 2, 页码 264-267

出版社

KARGER
DOI: 10.1159/000521530

关键词

Neonatology; Resuscitation; Heart rate; Electrocardiograms; Technology

资金

  1. Innovate UK [MRC-MC_PC_15012]

向作者/读者索取更多资源

This study evaluated the application time of pre-set ECG chest electrodes in newborns. The results showed that these electrodes can rapidly provide heart rate information without detachment or compromising the integrity of the skin.
Background: International newborn resuscitation guidelines recommend electrocardiogram (ECG) heart rate (HR) monitoring at birth. We evaluated the application time of pre-set ECG electrodes fixed to a polyethene patch allowing adhesive-free attachment to the wet skin of the newborn chest. Objectives: Using a three-electrode pre-set ECG patch configuration, application success was calculated using video analysis and measured at three time points, the time to (1) apply electrodes; (2) detect recognizable QRS complexes after application; and (3) display a HR after application. Method: A prospective observational study in two UK tertiary maternity units was undertaken with 71 newborns including 23 who required resuscitation. Results: The median (IQR) time for ECG patch application was 8 (6-10) seconds, detection of recognizable QRS complexes 8 (2-12) seconds, and time to output HR was 23 (15-37) seconds. Conclusion: Pre-set ECG chest electrodes allow rapid HR information at birth without electrode detachment or compromising skin integrity.