Disposable point-of-care portable perfusion phantom for quantitative DCE-MRI

期刊

MEDICAL PHYSICS
卷 49, 期 1, 页码 271-281

出版社

WILEY
DOI: 10.1002/mp.15372

关键词

abdomen; DCE-MRI; phantom; quantitative measurement

资金

  1. NCI [UG3CA232820]

向作者/读者索取更多资源

The DP4 phantom is reliable, portable, and capable of significantly improving the reproducibility of qDCE-MRI measurements.
Purpose To develop a disposable point-of-care portable perfusion phantom (DP4) and validate its clinical utility in a multi-institutional setting for quantitative dynamic contrast-enhanced magnetic resonance imaging (qDCE-MRI). Methods The DP4 phantom was designed for single-use and imaged concurrently with a human subject so that the phantom data can be utilized as the reference to detect errors in qDCE-MRI measurement of human tissues. The change of contrast-agent concentration in the phantom was measured using liquid chromatography-mass spectrometry. The repeatability of the contrast enhancement curve (CEC) was assessed with five phantoms in a single MRI scanner. Five healthy human subjects were recruited to evaluate the reproducibility of qDCE-MRI measurements. Each subject was imaged concurrently with the DP4 phantom at two institutes using three 3T MRI scanners from three different vendors. Pharmacokinetic (PK) parameters in the regions of liver, spleen, pancreas, and paravertebral muscle were calculated based on the Tofts model (TM), extended Tofts model (ETM), and shutter speed model (SSM). The reproducibility of each PK parameter over three measurements was evaluated with the intraclass correlation coefficient (ICC) and compared before and after DP4-based error correction. Results The contrast-agent concentration in the DP4 phantom was linearly increased over 10 min (0.17 mM/min, measurement accuracy: 96%) after injecting gadoteridol (100 mM) at a constant rate (0.24 ml/s, 4 ml). The repeatability of the CEC within the phantom was 0.997 when assessed by the ICC. The reproducibility of the volume transfer constant, K-trans, was the highest of the PK parameters regardless of the PK models. The ICCs of K-trans in the TM, ETM, and SSM before DP4-based error correction were 0.34, 0.39, and 0.72, respectively, while those increased to 0.93, 0.98, and 0.86, respectively, after correction. Conclusions The DP4 phantom is reliable, portable, and capable of significantly improving the reproducibility of qDCE-MRI measurements.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据