Radiation field and dose inhomogeneities using an X-ray cabinet in radiation biology research

期刊

MEDICAL PHYSICS
卷 48, 期 12, 页码 8140-8151

出版社

WILEY
DOI: 10.1002/mp.15297

关键词

53BP1; field inhomogeneity; radiation biology; radiation-induced foci; self-contained X-ray cabinet irradiator; thermoluminescence dosimetry; gamma H2AX

资金

  1. German Federal Ministry of Education and Research [02NUK035D]

向作者/读者索取更多资源

X-ray cabinets are user-friendly for biological radiation research, but field inhomogeneities and experimental setup components can significantly affect the delivered irradiation doses. Strict dosimetric monitoring is necessary for reliable studies.
Purpose: X-ray cabinets are replacing Cs-137/Co-60 sources in radiation biology research due to advantages in size, handling, and radiation protection. However, because of their different physical properties, X-ray cabinets are more susceptible to experimental influences than conventional sources. The aim of this study was to examine the variations related to the experimental setups typically used to investigate biological radiation effects with X-ray cabinets. Materials and methods: A combined approach of physical dose measurements by thermoluminescence dosimetry and detection of biological effects by quantification of gamma H2AX and 53BP1 foci was used to analyze field inhomogeneity and evaluate the influence of the components of the experimental setup. Results: Irradiation was performed using an X-ray tube (195 kV, 10 mA, 0.5-mm-thick copper filter, dose rate of 0.59 Gy/min). Thermoluminescence dosimetry revealed inhomogeneity and a dose decrease of up to 42.3% within the beam area (diameter 31.1 cm) compared to the dose at the center. This dose decrease was consistent with the observed decline in the number of radiation-induced foci by up to 55.9 %. Uniform dose distribution was measured after reducing the size of the radiation field (diameter 12.5 cm). However, when using 15-ml test tubes placed at different positions within this field, the dose decreased by up to 17% in comparison to the central position. Analysis of foci number revealed significant differences between the tubes for gamma H2AX (1 h) and 53BP1 (4 h) at different time points after irradiation. Neither removal of some tubes nor of the caps improved the dose decrease significantly. By contrast, when using 1.5-ml tubes, dose differences were less than 4%, and no significant differences in foci number were detected. Conclusion: X-ray cabinets are user-friendly irradiation units for investigating biological radiation effects. However, field inhomogeneities and experimental setup components considerably affect the delivered irradiation doses. For this reason, strict dosimetric monitoring of experimental irradiation setups is mandatory for reliable studies.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据