4.4 Article

Theoretical design of an absorption hologram-based sensor for dose quantification in daylight photodynamic therapy

出版社

OPTICAL SOC AMER
DOI: 10.1364/JOSAA.441684

关键词

-

类别

向作者/读者索取更多资源

This paper presents a novel configuration for a D-PDT dose sensor that can be placed adjacent to the treatment site to accurately quantify the light dose received. The theoretical modeling and experimental results show that the initial absorption and grating thickness have a significant impact on the magnitude and rate of the sensor response.
Daylight photodynamic therapy (D-PDT) is an effective and almost painless treatment for many skin conditions, where successful treatment relies on daylight activation of a topical photosensitizer. Optimization of D-PDT requires accurate assessment of light dose received. There is a requirement for a small-area sensor that can be placed adjacent to the treatment site to facilitate accurate dose quantification. Here, a novel, to the best of our knowledge, configuration for a D-PDT dose sensor, consisting of a holographic absorption grating fabricated in a photosensitive film, is presented. Theoretical modeling of the sensor's response (i.e., change in grating diffraction efficiency due to change in grating absorption modulation, alpha(1), on exposure to daylight) was conducted using Kogelnik's coupled-wave theory. The influence of the different grating parameters (initial film absorption, thickness, spatial frequency, and reconstruction wavelength) on the sensor response was examined and revealed that the initial absorption and grating thickness values have a large impact on both the magnitude and rate of the D-PDT sensor response. The optimum design for an absorption grating-based D-PDT sensor is described. (C) 2021 Optical Society of America

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据