The study evaluated the mechanical, micro-mechanical, and physical properties of Reva Medical Fantom Encore Bioresorbable Scaffolds subjected to thermally-accelerated degradation. Results showed an initial increase in radial strength and stiffness, followed by a decrease in ductility and load-bearing capacity with continued degradation. The scaffolds exhibited increased crystallinity throughout the degradation process, with minimal changes in nano-mechanical properties.
This study presents a comprehensive evaluation of the mechanical, micro-mechanical and physical properties of Reva Medical Fantom Encore Bioresorbable Scaffolds (BRS) subjected to a thermally-accelerated degradation protocol. The Fantom Encore BRS were immersed in phosphate buffered saline solution at 50 degrees C for 112 days with radial compression testing, nanoindentation, differential scanning calorimetry, gel permeation chromatography and mass loss characterisation performed at consecutive time points. In the initial stages of degradation (Days 0-21), the Fantom Encore BRS showed increases in radial strength and stiffness, despite a substantial reduction in in molecular weight, with a slight increase in the melt temperature also observed. In the second phase (Days 35-54), the radial strength of the BRS samples were maintained despite a continued loss in molecular weight. However, during this phase, the ductility of the stent showed a reduction, with stent fracture occurring earlier in the crimp process and with lower amounts of plastic deformation evident under visual examination post-fracture. In the final phase (Days 63-112), the load-bearing capacity of the Fantom Encore BRS showed continued reduction, with decreases in radial stiffness and strength, and drastic reduction in the work-to-fracture of the devices. Throughout each phase, there was a steady increase in the relative crystallinity, with limited mass loss until day 112 and only minor changes in glass transition and melt temperatures. Limited changes were observed in nano-mechanical properties, with measured local elastic moduli and hardness values remaining largely similar throughout degradation. Given that the thermally-accelerated in vitro conditions represented a four-fold acceleration of physiological conditions, these results suggest that the BRS scaffolds could exhibit substantially brittle behaviour after similar to one year of implantation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据