4.5 Article

Personalised nitinol stent for focal plaques: Design and evaluation

期刊

JOURNAL OF BIOMECHANICS
卷 130, 期 -, 页码 -

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.jbiomech.2021.110873

关键词

Femoropopliteal artery; Finite element; Focal plaque; Nitinol stent; Personalised design

资金

  1. EPSRC UK [EP/R001650/1]
  2. EPSRC [EP/R001650/1] Funding Source: UKRI

向作者/读者索取更多资源

The personalised nitinol stent designs show significant improvements in lumen gain, reduction of stress levels in the media layer, and enhancement of lumen shape compared to the commercial stent design. This pilot study highlights the potential of personalised stents in delivering desirable treatment outcomes.
The purpose of this study is to develop personalised nitinol stents for arteries with one and two opposite focal plaques. Novel designs are evaluated through comparison with a commercial stent design, in terms of lumen gain and shape as well as stress levels in the media layer after stenting. Methods: Personalised stents are developed for arteries with one and two opposite focal plaques, based on medical imaging of patients and computer simulations. In silico analysis is then carried out for assessment of stent performance in the diseased arteries. Results: Personalised designs significantly increase the lumen gain, reduce the stresses in the media layer, and improve the lumen shape compared to the commercial nitinol stent. Conclusion: The personalised designs show outstanding performance compared to the commercial stent. Significance: This pilot study proves that personalised nitinol stents are able to deliver desirable treatment outcomes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据