Characterization and commissioning of a Leksell Gamma Knife ICON system for framed and frameless stereotactic radiosurgery

期刊

出版社

WILEY
DOI: 10.1002/acm2.13475

关键词

CBCT; end-to-end testing; film dosimetry; Gamma Knife IU; SRS

向作者/读者索取更多资源

This study mainly evaluated the dosimetry and targeting accuracy of the Leksell Gamma Knife Icon unit (IU) for both invasive frame and frameless treatment workflows. Measurements and tests were conducted using EBT3 film, CBCT and CT-based methods, and a rigid phantom. The results showed that the IU treatment deliveries had reliable dosimetric and geometric accuracy, and the commissioning of IU for clinical use was confirmed.
Purpose The Leksell Gamma Knife Icon unit (IU) was introduced recently as an upgrade to the Perfexion unit (PU) at our Gamma Knife practice. In the current study, we sought mainly to characterize dosimetry and targeting accuracy of the IU treatment deliveries using both invasive frame and frameless treatment workflows. Methods Relative output factors were measured by delivering single-shot 4, 8 and 16 mm radiation profiles in the manufacturer's acrylonitrile butadiene styrene spherical phantom in coronal and sagittal planes using EBT3 film. Resultant dosimetry was compared with the manufacturer's dose calculation and derived output factors were compared with the manufacturer's published value. Geometric consistency of stereotactic coordinates based on cone-beam computed tomography (CBCT) versus the traditional conventional-CT-based method was characterized using a rigid phantom containing nine fiducial indicators over four separate trials. End-to-end (E2E) testing using EBT3 film was designed to evaluate both dosimetric and geometric accuracy for hypothetical framed and frameless workflows. Results Relative output factors as measured by the manufacturer were independently confirmed using EBT3 film measurements to within 2%. The mean 3D radial discrepancy in stereotactic space between CBCT and CT-based definition over the sampled locations in our rigid geometry phantom was demonstrated to be between 0.40 mm and 0.56 mm over the set of trials, larger than prior reported values. E2E performed in 2D demonstrates sub-mm (and typically < 0.5 mm) accuracy for framed and frameless workflows; geometric accuracy of framed treatments using CBCT-defined stereotactic coordinates was shown to be slightly improved in comparison with those defined using conventional CT. Furthermore, in phantom, frameless workflows exhibited better accuracy than framed workflows for fractionated treatments, despite large magnitudes of introduced interfraction setup error. Accuracy of dosimetric delivery was confirmed in terms of qualitative comparisons of dose profiles and in terms of 2D gamma pass rates based on 1%/1 mm criteria. Conclusion The IU was commissioned for clinical use of frameless and framed treatment protocols. The present study outlines an extensive E2E methodology for confirmation of dosimetric and geometric treatment accuracy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据