Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers

出版社

SPRINGER
DOI: 10.1007/s00417-022-05579-9

关键词

Repeatability; Agreement; Anterion; Anterior segment biometry; Axial length; Optical coherence tomographer

资金

  1. Wenzhou Key Team of Scientific and Technological Innovation [C20170002]
  2. Wenzhou Public Welfare Science and Technology Projects [Y20170192]
  3. Young Talents Programme of Zhejiang Medical and Health Science and Technology Project [2019RC223]
  4. Engineering Development Project of Ophthalmology and Optometry [GCKF201601]
  5. Nature and Science Foundation of China [81570869]
  6. Zhejiang Provincial Foundation of China for Distinguished Young Talents in Medicine and Health [2010QNA018]

向作者/读者索取更多资源

This study investigated the repeatability of Anterion and compared its agreement with other three optical biometers in terms of ocular biometric measurements and predicted IOL powers. The results showed that Anterion exhibited good repeatability for most parameters and had good agreement with other biometers, except for the predicted IOL powers.
Purpose To investigate the repeatability of Anterion and compare the agreement of ocular biometric measurements and predicted intraocular lens (IOL) powers with other three optical biometers. Methods Flat keratometry (Kf), steep keratometry (Ks), J0 and 45 vectors, central cornea thickness (CCT), anterior chamber depth (ACD), and axial length (AL) from the Anterion, IOLMaster 700, Lenstar LS 900, and OA-2000 were recorded. The IOL powers were calculated with the Hoffer Q, Holladay 1, SRK/T, and Haigis formulas. The repeatability was evaluated using the within-subject standard deviation (Sw), repeatability coefficient (RC), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Inter-device agreement between the four biometers was assessed with the 95% limits of agreement. Results In total, 101 right eyes of 101 participants were enrolled. The Anterion showed good repeatability for all the included biometric parameters with all the CoV <= 0.30% and ICC >= 0.930 except for J45 with moderate repeatability (ICC was 0.849). Good agreement was found among the four devices for Kf, Ks, J0, J45, ACD, and AL. Generally, wide 95% LoA was found for the predicted IOL powers with the four IOL calculation formulas between the four devices. Conclusions The Anterion showed good repeatability of biometric measurements for most parameters. Good agreement among the four optical biometers was achieved for all the parameters except for CCT and the predicted IOL power. The AL values exhibited the best repeatability with Anterion and the best agreement among the biometers in our study.